Control Test Tablets

Control Test Tablets for verifying the full functionality of the RESPIROMETRIC Sensor System by simulating an oxygen demand experiment similar to the BOD method

Product Code: A00000136

Description

Features:
• Reaction Dynamic: When dissolved in distilled water, the tablet components consume a precise, stoichiometrically predetermined volume of oxygen gas from the bottle’s headspace.
• Resulting Effect: This chemical consumption generates a highly reproducible negative pressure gradient inside the airtight system.
• Validation Purpose: The pressure change tests the electronic manometric sensors, check-valves, system seals, stirring consistency, and incubator temperature regulation simultaneously over a 5-day cycle.

Additional information

Required Liquid Volume

157 ml of pure distilled water per bottle

Incubator Temperature

20.0 °C ± 0.5 °C

Measurement Scale Setup

600 mg/l (ppm) full scale

CO₂ Absorber Modification

The alkali holder must be left empty (Do not add KOH/NaOH crystals or traps)

Experiment Duration

5 Days

Sampling Interval

Every 8 hours (standard logging via software)

Expected Nominal Value

314 ± 30 mg/l (or 293 ± 30 mg/l depending on the specific application version/lot calibration curve)